Exploration of the impact of body mass index on radiotherapy setup error in nasopharyngeal carcinoma patients
Background and purpose: Radiotherapy is the main treatment for nasopharyngeal carcinoma. This study aimed to investigate the potential association between body mass index (BMI) and setup error during daily delivery of radiotherapy in the patients with nasopharyngeal carcinoma. Methods: This study enrolled 50 patients with nasopharyngeal carcinoma
who were treated with radical radiation therapy in Fudan University Shanghai Cancer Center from January 2017 to June 2017. The body weight was measured before radiation and 4 weeks post-radiation. Cone beam CT was performed at the same time point to obtain position data of the patients
including the patient’s head pin (LNG)
lateral (LAT)
dorsoventral (VRT) directions. Based on these analyses
we evaluated the accuracy of patient setup
as well as the adequacy of margins during target volume contouring. Results: The average body weights of patients before radiotherapy and 4 weeks post-radiotherapy were (53.27±3.79) and (49.37±3.29) kg
respectively. There was significant decrease of body weight during radiotherapy (P0.05). System error at LNG
LAT and VRT detections were 2.6
2.0 and 3.2 mm
respectively. The random error of LNG
LAT and VRT were 2.9
2.0 and 3.8 mm
respectively. BMI 24 was chosen to be the cutting point. In the subgroup with BMI≥24 kg/m2
the system errors at three-dimensional direction and horizontal rotation position were (2.80±0.35)
(2.30±0.17) and (3.90±0.42) mm
respectively. In the subgroup with BMI24 kg/m2
the system errors at three directions were (2.30±0.12)
(1.90±0.22) and (3.00±0.67) mm
respectively. There was significant difference in LNG
LAT and VRT between the two subgroups (P0.05). The correlation analysis revealed no significant association between setup values and BMI (P0.05). In other words
the setup error was higher in the subgroups with BMI≥24 kg/ m2. Decreasing BMI shifted the central point of patient setup towards ventral side and rotated the patient setup point reversely. Conclusion: There was no significant association between BMI and setup error before radiotherapy in patients with nasopharyngeal carcinoma. However
weight loss during radiotherapy influenced accuracy of patient setup. To ensure the accuracy of radiation
certain corrections have to be made during radiotherapy.
Investigation of patient characteristics associated with setup errors in intensity-modulated radiotherapy after breast-conserving surgery
The diagnostic value of zero echo time magnetic resonance imaging for skull base bone invasion in nasopharyngeal carcinoma
Impact of radiation-induced lymphopenia on prognosis in patients with limited-stage small cell lung cancer: a cohort study
Derepression of retrotransposable elements in the development of radiation-induced late effects: advancements and perspective
A comparative analysis of the long-term efficacy between partial hepatectomy and transcatheter arterial chemoembolization in patients with liver metastasis from nasopharyngeal carcinoma
Related Author
尹志海
汪隽琦
孟怡然
许 青
杨昭志
Jiahao LIN
Meimei FENG
Kongqi LIN
Related Institution
复旦大学附属肿瘤医院放射治疗中心,复旦大学上海医学院肿瘤学系
Department of Radiation Oncology, Fujian Cancer Hospital, Clinical Oncology School of Fujian Medical University
Department of Radiology, Fujian Cancer Hospital, Clinical Oncology School of Fujian Medical University
Department of Thoracic Radiation Oncology, Fudan University Shanghai Cancer Center, Department of Oncology, Shanghai Medical College, Fudan University
Department of Oncology, Taiyuan Central Hospital/The Ninth Clinical Medical College of Shanxi Medical University