甲状腺CT辐射剂量、图像质量与女性体成分的相关性研究

沈茜刚1(), 林呈祥2(), 彭卫军1, 顾雅佳1()   

  1. 1.复旦大学附属肿瘤医院放射诊断科,复旦大学上海医学院肿瘤学系,上海 200032
    2.复旦大学附属儿科医院放射科,国家儿童医学中心,上海 201102
  • 收稿日期:2024-08-14 修回日期:2024-12-18 出版日期:2025-04-30 发布日期:2025-05-16
  • 通信作者: 顾雅佳(ORCID: 0000-0002-6130-859X),主任医师,教授,复旦大学附属肿瘤医院放射诊断科主任。
  • 作者简介:沈茜刚(ORCID: 0000-0001-6460-3474),副主任技师;
    共同第一作者:林呈祥(ORCID: 0009-0008-3902-329X),主管技师。

A study on correlation between thyroid CT radiation dose, image quality and female body composition

SHEN Xigang1(), LIN Chengxiang2(), PENG Weijun1, GU Yajia1()   

  1. 1. Department of Radiology, Fudan University Shanghai Cancer Center; Department of Oncology, Shanghai Medical College, Fudan University, Shanghai 200032, China
    2. Department of Radiology, Children's Hospital of Fudan University, National Children's Medical Center, Fudan University, Shanghai 201102, China
  • Received:2024-08-14 Revised:2024-12-18 Published:2025-04-30 Online:2025-05-16
  • Contact: GU Yajia

摘要:

背景与目的:计算机体层成像(computed tomography,CT)检查已成为有效剂量的主要来源之一。本研究通过分析甲状腺CT的辐射剂量、图像质量和女性体成分之间的相关性,为今后更好地控制辐射剂量和图像质量提供数据基础。方法:回顾性分析2015年1月—2019年12月在复旦大学附属肿瘤医院进行甲状腺CT检查的患者。收集甲状腺实质均匀区域的CT值和标准差值,相同感兴趣区(region of interest,ROI)大小测量邻近皮下脂肪组织的CT值和标准差值。所有数据需测3次取其均值,记录容积CT剂量指数(CT dose index of volume,CTDIvol)、剂量长度乘积(dose length product,DLP)并计算有效剂量值和各部位的信噪比(signal-to-noise ratio,SNR)、对比噪声比(contrast-to-noise ratio,CNR)。所有患者均接受骨密度检查,记录患者的身高、体重、体重指数(body mass index,BMI)、体表面积(body surface area,BSA)、脊柱与大腿肌肉百分比、脊柱与大腿脂肪百分比的数值。结果:104例患者纳入本研究。甲状腺CT图像的SNR和脊柱肌肉百分比(r=0.284,P=0.004)、CNR和脊柱脂肪百分比(r=0.197,P=0.045)、甲状腺CT图像的有效剂量和患者的年龄、体重、BMI、大腿脂肪百分比、BSA之间(r=0.221、0.247、0.260、0.262和0.222,P=0.024、0.011、0.008、0.007和0.024)均呈正相关。SNR和年龄、体重、BMI及脊柱脂肪百分比(r=-0.292、-0.198、-0.207、-0.284;P=0.003、0.044、0.035和0.004),CNR和脊柱肌肉百分比(r=-0.197,P=0.045),有效剂量和大腿肌肉百分比(r=-0.262,P=0.007)之间均呈负相关。结论:甲状腺CT的辐射剂量与女性体成分有相关性,图像质量与女性体成分的相关关系为今后更好地控制辐射剂量和提高图像质量提供了数据基础。

关键词: 辐射剂量, 体成分, 有效剂量, 信噪比, 对比噪声比

Abstract:

Background and purpose: With the increasing use of medical imaging examinations, Computed tomography (CT) scans have become one of the primary sources of effective dose. This study analyzed the correlation between radiation dose, image quality and body composition in thyroid CT to provide a data foundation for better control of radiation dose and image quality in the future. Methods: A retrospective analysis was conducted on patients who underwent thyroid CT scans at Fudan University Shanghai Cancer Center from January 2015 to December 2019. CT values and standard deviations (SD) of the homogeneous parenchyma of the thyroid were collected, and CT values and SD values of adjacent subcutaneous fat tissue were measured with the same region of interest (ROI) size. All data were measured three times to obtain the average value, and the volume CT dose index (CTDIvol) and dose-length product (DLP) were recorded. Effective dose values and signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) for various parts were calculated. All patients underwent bone density examinations, and patient height, weight, body mass index (BMI), body surface area (BSA), and the percentage of muscle and fat in the spine and thighs were recorded. Results: A total of 104 patients were enrolled in this study. The SNR of thyroid CT images and the percentage of spinal muscle (r=0.284, P=0.004), CNR of thyroid CT images and the percentage of spinal fat (r=0.197, P=0.045), and the effective dose of thyroid CT images and age, weight, BMI, thigh fat percentage, and body surface area (r=0.221, 0.247, 0.260, 0.262, and 0.222; P=0.024, 0.011, 0.008, 0.007 and 0.024) were positively correlated. While, SNR and age, weight, BMI, and the percentage of spinal fat (r=-0.292, -0.198, -0.207, and -0.284; P=0.003, 0.044, 0.035, 0.004), CNR and the percentage of spinal muscle (r=-0.197, P=0.045), and the effective dose and the percentage of thigh muscle (r=-0.262, P=0.007) were negatively correlated. Conclusion: The radiation dose in thyroid CT is correlated with female body composition. The image quality is related to female body composition, providing a data foundation for better control of radiation dose and image quality in the future.

Key words: Radiation dose, Body composition, Effective dose, Signal-to-noise ratio, Contrast-to-noise ratio