Establishment of daily quality assurance program and analysis of dosimetric characteristics’ longterm stability for mobile intraoperative radiation therapy accelerator

WANG Jun-qi,MA Jin-li,HU Wei-gang,SUN Li-ning   

  1. Department of Radiation Oncology, Fudan University Shanghai Cancer Center,Department of Oncology, Shanghai Medical College, Fudan University, Shanghai 200032, China
  • Online:2014-01-30 Published:2014-02-14
  • Contact: MA Jin-li E-mail: jinlima_c@yahoo.com

Abstract: Background and purpose: Currently, intraoperative radiation therapy (IORT) has become the adjuvant therapy of cancer. The study was to establish the daily quality assurance (QA) program and analyze dosimetric characteristics’ long-term stability for mobile IORT accelerator. Methods: The QA program of this study included two parts: safety and functionality and energy and output. The two years’ QA datasets were acquired and analyzed to investigate the stability of energy and output. Results: All safety and functionality tests passed on a daily basis. The energy index was (0.666±0.015)mm, (0.839±0.009)mm, (0.781±0.010)mm, (0.724±0.009)mm and the output dose error was (0.511 ± 0.671)%, (0.278 ± 0.516)%, (0.368 ± 0.532)%, (0.382 ± 0.912)% for all energy, respectively. There was no significant time trend in the dosimetric characteristics. Conclusion: The daily QA program is suitable for mobile IORT accelerator. The long-term stability is acceptable for IORT in clinical use.

Key words: Intraoperative radiation therapy, Quality assurance, Long-term stability